Finite element analysis of the mechanical strength of a new hip Spacer

文献详情

资源类型:
Pubmed体系:
机构: [1]The First Clinical Medical School, Guangzhou University of Chinese Medicine, Jichang Road 12#, District Baiyun, Guangzhou, Guangdong, China. [2]Department of Orthopaedics, The First Affiliated Hospital of Guangzhou University of Chinese Medicine, Jichang Road 16#, District 22 Baiyun, Guangzhou, 510405, Guangdong, China.
出处:

关键词: Finite element Periprosthetic joint infection Bone cement Spacer

摘要:
At present, the influence of the internal metallic endoskeleton of Spacer on the biomechanical strength of Spacer remains unclear. The aim of this study was to analyze the mechanical stability of a novel Spacer applying a annular skeleton that mimics the structure of trabecular bone using finite element methods.The femur models of three healthy individuals and skeletonless Spacer, K-Spacer, and AD-Spacer were assembled to create 15 3D models. Finite element analysis was performed in an Ansys Bench2022R1. Biomechanical parameters such as stress and strain of the Spacer, internal skeleton and femur were evaluated under three loads, which were applied with the maximum force borne by the hip joint (2100 N), standing on one leg (700 N), and standing on two legs (350 N). The mechanical properties of the new hip Spacer were evaluated.The stresses on the medial and lateral surfaces of the AD-Spacer and K-Spacer were smaller than the stresses in the state without skeletal support. The maximum stresses on the medial and lateral surfaces of the AD-Spacer were smaller than those of the inserted K-Spacer, and the difference gradually increased with the increase of force intensity. When the skeleton diameter was increased from 3 to 4 mm, the stresses in the medial and lateral sides of the AD-Spacer and K-Spacer necks decreased. The stress of both skeletons was concentrated at the neck, but the stress of the annular skeleton was evenly distributed on the medial and lateral sides of the skeleton. The mean stress in the proximal femur was higher in femurs with K-Spacer than in femurs with AD-Spacer.AD-Spacer has lower stress and higher load-bearing capacity than K-Spacer, and the advantages of AD-Spacer are more obvious under the maximum load state of hip joint.© 2023. The Author(s).

基金:
语种:
PubmedID:
中科院(CAS)分区:
出版当年[2022]版:
大类 | 3 区 医学
小类 | 4 区 骨科 4 区 风湿病学
最新[2025]版:
大类 | 3 区 医学
小类 | 3 区 骨科 3 区 风湿病学
第一作者:
第一作者机构: [1]The First Clinical Medical School, Guangzhou University of Chinese Medicine, Jichang Road 12#, District Baiyun, Guangzhou, Guangdong, China.
通讯作者:
推荐引用方式(GB/T 7714):
APA:
MLA:

资源点击量:2018 今日访问量:0 总访问量:645 更新日期:2024-07-01 建议使用谷歌、火狐浏览器 常见问题

版权所有©2020 广东省中医院 技术支持:重庆聚合科技有限公司 地址:广州市越秀区大德路111号