Performances evaluation of four systems for homocysteine determination by LC-MS/MS reference method

文献详情

资源类型:
WOS体系:

收录情况: ◇ SCIE

机构: [a]Department of Laboratory Medicine, The Second Affiliated Hospital of Guangzhou University of Chinese Medicine, Guangzhou, China [b]Second Clinical Medical College, Guangzhou University of Chinese Medicine, Guangzhou, China [c]Southern Medical University, Guangzhou, China
出处:
ISSN:

关键词: Accuracy Comparability study ID-LC-MS/MS Performance evaluation Precision Reference method Serum homocysteine

摘要:
Background: The aim of this study is to verify the analytical performance of four homocysteine detection systems made in China and to explore the comparability of homocysteine detection systems by isotope dilution liquid chro-matography-tandem mass spectrometry (ID-LC-MS/MS) reference method. Methods: The intra-batch precision, inter-batch precision, accuracy, and linear range of four homocysteine detec-tion systems were evaluated. The ID-LC-MS/MS reference method was used to evaluate the comparability and ac-curacy of fresh frozen serum samples in four different detection systems of homocysteine. The ID-LC-MS/MS ref-erence method is used to assign samples as calibrators to calibrate each system. The variation and deviation of fresh serum samples between different systems before and after calibration were compared. Results: The intra-batch imprecision of the four detection systems was less than 5%, and the coefficient of varia-tion of inter-batch imprecision was less than 6.7%. The precision met the clinical requirements. Before calibra-tion, the results measured by detection system 2 are consistent with the ID-LC-MS/MS reference method, which meets the requirements of accuracy verification. The regression equation of R ≥ 0.975 in the regression equation of linear analysis of the four systems, the linearity of the four detection systems is good in the range of evaluation concentration, and all of them can meet the declared linear range. The absolute average bias of fresh serum mea-sured by the four detection systems after calibration decreased from 3.76 μmol/L, 0.96 μmol/L, 1.30 μmol/L, -1.56 μmol/L to 0.31 μmol/L, 0.28 μmol/L, 0.4 μmol/L, 0.40 μmol/L, respectively. The relative average bias de-creased from 22.6%, 7.50%, 11.0% and -8.50% to 1.98%, 1.78%, 2.59%, 2.34%, respectively. After calibration, the slope and intercept of the regression curve of the fresh serum measured by the four detection systems and the reference method are closer to 1 and 0 than before calibration. Conclusions: The precision, reference interval, and linear evaluation of the four detection systems are good. The ID-LC-MS/MS reference method assigning fresh frozen serum samples as calibrators can improve the accuracy and comparability of the results of different detection systems. © 2021 Verlag Klinisches Labor GmbH. All rights reserved.

基金:

基金编号: 2019YFF0216505

语种:
WOS:
PubmedID:
中科院(CAS)分区:
出版当年[2020]版:
大类 | 4 区 医学
小类 | 4 区 医学实验技术
最新[2025]版:
大类 | 4 区 医学
小类 | 4 区 医学实验技术
JCR分区:
出版当年[2019]版:
Q4 MEDICAL LABORATORY TECHNOLOGY
最新[2023]版:
Q4 MEDICAL LABORATORY TECHNOLOGY

影响因子: 最新[2023版] 最新五年平均 出版当年[2019版] 出版当年五年平均 出版前一年[2018版] 出版后一年[2020版]

第一作者:
第一作者机构: [a]Department of Laboratory Medicine, The Second Affiliated Hospital of Guangzhou University of Chinese Medicine, Guangzhou, China
通讯作者:
通讯机构: [a]Department of Laboratory Medicine, The Second Affiliated Hospital of Guangzhou University of Chinese Medicine, Guangzhou, China
推荐引用方式(GB/T 7714):
APA:
MLA:

资源点击量:2018 今日访问量:0 总访问量:645 更新日期:2024-07-01 建议使用谷歌、火狐浏览器 常见问题

版权所有©2020 广东省中医院 技术支持:重庆聚合科技有限公司 地址:广州市越秀区大德路111号