Interference of Thalassemias on Hemoglobin A1c Measurement by Ion-Exchange High-Performance Liquid Chromatography Method Tosoh HLC-723 G8

文献详情

资源类型:
WOS体系:

收录情况: ◇ SCIE

机构: [1]Guangzhou Univ Chinese Med, Dept Lab Sci, Affiliated Hosp 2, Dade Rd 111, Guangzhou 510120, Guangdong, Peoples R China
出处:
ISSN:

关键词: Hemoglobin A(1c) (HbA(1c)) thalassemia ion-exchange high-performance liquid chromatography (IE-HPLC)

摘要:
Background: The presence of hemoglobinopathies could interfere with some assays for Hemoglobin A(1c) (HbA(1c)) measurement; therefore, the effect of thalassemia on ion-exchange high-performance liquid chromatography (IEHPLC) method Tosoh HLC-723 G8 (Tosoh G8) was evaluated. Methods: A total of 43 normal controls and 101 thalassemia patients were quantified by Premier Hb9210 and Tosoh G8 (variant-mode) systems. At the same time, 7 normal controls and 8 thalassemia patients were confirmed by the International Federation of Clinical Chemistry and Laboratory Medicine (IFCC) reference method for verification. Results: For normal controls, the HbA(1c) values of Tosoh G8 system (y) showed great correlation and agreement with those from Premier Hb9210 (x) (y = 0.9688x + 0.2151, r = 0.9951; mean difference 0.02 +/- 0.30%), and no significant relative bias above 7% was observed; the HbA(1c) values obtained by Tosoh G8 were consistent with the IFCC targets (relative bias < +/- 6%) in all of the samples. However, for thalassemia, the correlation between Tosoh G8 (y) and Premier Hb9210 (x) became relatively low (y = 0.8079x + 1.2897, r = 0.7780); the HbA ie values of 91.1% of the samples (92/101) obtained by Tosoh G8 were higher than those by Premier Hb9210 (mean difference 0.33 +/- 0.48%) and a significant positive bias above 7% was noticed in 43.3% (45/101) thalassemia patients; when compared with the IFCC targets, the 87.5% (7/8) relative bias was > +/- 6%. Conclusions: Thalassemia could directly affect the measurement of HbA(1c) using the IE-HPLC method Tosoh G8 and the clinical laboratorial staff should pay close attention.

基金:

基金编号: 81572088 2015A030313340 2014KT1593 2012BAI37B01 2012ZX09303009-003 201007005 2013BAI02B04

语种:
WOS:
PubmedID:
中科院(CAS)分区:
出版当年[2017]版:
大类 | 4 区 医学
小类 | 4 区 医学实验技术
最新[2025]版:
大类 | 4 区 医学
小类 | 4 区 医学实验技术
JCR分区:
出版当年[2016]版:
Q4 MEDICAL LABORATORY TECHNOLOGY
最新[2023]版:
Q4 MEDICAL LABORATORY TECHNOLOGY

影响因子: 最新[2023版] 最新五年平均 出版当年[2016版] 出版当年五年平均 出版前一年[2015版] 出版后一年[2017版]

第一作者:
第一作者机构: [1]Guangzhou Univ Chinese Med, Dept Lab Sci, Affiliated Hosp 2, Dade Rd 111, Guangzhou 510120, Guangdong, Peoples R China
推荐引用方式(GB/T 7714):
APA:
MLA:

资源点击量:2020 今日访问量:0 总访问量:646 更新日期:2024-07-01 建议使用谷歌、火狐浏览器 常见问题

版权所有©2020 广东省中医院 技术支持:重庆聚合科技有限公司 地址:广州市越秀区大德路111号