Performance evaluation of a new high throughput Mindray BS-2000M1 clinical chemistry system

文献详情

资源类型:
WOS体系:
Pubmed体系:

收录情况: ◇ SCIE

机构: [1]Department of Laboratory Medicine, Guangdong Provincial Hospital of Chinese Medicine, Guangzhou, China [2]Department of Stomatology, Guangzhou Women and Children's Medical Center, Guangzhou, China [3]Department of Laboratory Medicine, The First Affiliated Hospital, Sun Yat-Sen University, Guangzhou, China
出处:
ISSN:

关键词: Evaluation BS-2000M1 Cobas 8000 Technical performance Analytical performance

摘要:
Objective: To evaluate the photometry technical and analytical performance of a newly launched Mindray BS-2000M1 clinical chemistry system (BS-2000M1). Design and Methods: The photometric parameters were evaluated according to the China Food and Drug Administration (CFDA) Automatic Chemistry Analyzer Guideline. The precision, accuracy, linearity and interference were evaluated according to CLSI protocols EP5-A2, EP9-A2, EP6-A and EP7-A2 respectively. The trueness verification on Ca2+, Mg2+, P- and Cl- was conducted by comparing with the reference methods using fresh samples. Results: The photometer accuracy, precision, linearity, stability and stray light at 340 nm were acceptable. The within-run coefficients of variation (CVs) ranged from 0.16% to 2.13% and the total CV ranged from 0.64% to 4.12%. A good correlation (R > 0.95) of method comparison between BS-2000M1 and our reference system was observed for most of the parameters tested with exception of Ca2+ (R = 0.85), Mg2+ (R = 0.71), P- (R = 0.96, Slope = 0.88), Cl- (R = 0.93), and ASO (R = 0.94, Slope = 0.93, intercept = -8.81). The trueness verification on Ca2+, Mg2+, P- and Cl- showed acceptable results on both BS-2000M1 and our reference systems. Linearity study showed acceptable linearity range for all parameters. Significant interferences occurred for some evaluated parameters, but were identical to the manufacturer statement. Conclusions: Mindray BS-2000M1 achieved the desirable photometry technical and analytical performance, and is therefore suitable for its utilization in modern clinical laboratories. (C) 2014 The Canadian Society of Clinical Chemists. Published by Elsevier Inc. All rights reserved.

语种:
被引次数:
WOS:
PubmedID:
中科院(CAS)分区:
出版当年[2013]版:
大类 | 3 区 医学
小类 | 3 区 医学实验技术
最新[2025]版:
大类 | 3 区 医学
小类 | 3 区 医学实验技术
JCR分区:
出版当年[2012]版:
Q1 MEDICAL LABORATORY TECHNOLOGY
最新[2023]版:
Q2 MEDICAL LABORATORY TECHNOLOGY

影响因子: 最新[2023版] 最新五年平均 出版当年[2012版] 出版当年五年平均 出版前一年[2011版] 出版后一年[2013版]

第一作者:
第一作者机构: [1]Department of Laboratory Medicine, Guangdong Provincial Hospital of Chinese Medicine, Guangzhou, China
通讯作者:
推荐引用方式(GB/T 7714):
APA:
MLA:

资源点击量:2018 今日访问量:0 总访问量:645 更新日期:2024-07-01 建议使用谷歌、火狐浏览器 常见问题

版权所有©2020 广东省中医院 技术支持:重庆聚合科技有限公司 地址:广州市越秀区大德路111号