Liquid chromatography-tandem mass spectrometry analysis of 17-hydroxyprogesterone in dried blood spots revealed matrix effect on immunoassay

文献详情

资源类型:
WOS体系:
Pubmed体系:

收录情况: ◇ SCIE ◇ EI

机构: [1]Department of Laboratory Medicine, The Second Affiliated Hospital of Guangzhou University of Chinese Medicine, Guangzhou, Guangdong, China [2]New York State Department of Health, Wadsworth Center, Empire State Plaza, Albany, NY, USA [3]College of Arts and Sciences, University at Albany, State University of New York, Albany, NY, USA
出处:
ISSN:

关键词: Newborn screening Congenital adrenal hyperplasia 17-hydroxyprogesterone Cortisol Mass spectrometry

摘要:
Immunoassays for measuring 17-hydroxyprogesterone (17-OHP) produce high rates of false positives that impact the identification of congenital adrenal hyperplasia (CAH) in neonates. A confirmatory test with high analytical specificity employing liquid chromatography-tandem mass spectrometry (LC-MS/MS) methodology is needed in newborn screening for CAH. 17-OHP and cortisol were extracted from dried blood spot (DBS) samples, resolved on a C18 column, and measured using tandem mass spectrometry. The results were compared with those determined using the AutoDELFIA immunoassay. The LC-MS/MS method had a limit of quantitation of 10.0 and 5.0ng/mL for 17-OHP and cortisol, respectively. The method characteristics showed coefficient variation (%CV) 11.9% for both 17-OHP and cortisol, recoveries ranging from 83.1 to 101.5% for 17-OHP and from 95.1 to 102.8% for cortisol, and linearity with R-2=0.9994 for 17-OHP and R-2=0.9996 for cortisol, clinical sensitivity of 100.0% and a specificity of 96.4% as obtained by receiver operating characteristic analysis on 45 patient samples when 17-OHP >39.1ng/mL was selected as the cutoff value. Comparison between the LC-MS/MS and the AutoDELFIA immunoassay methods revealed a poor correlation for patient DBS samples (R-2=0.6784); however, an excellent correlation was obtained for QC and proficiency test (PT) DBS samples (R-2=0.9797). The LC-MS/MS method produced reliable results for 17-OHP and cortisol for the diagnosis of CAH. The AutoDELFIA immunoassay appears to be subject to matrix effects in the analysis for 17-OHP in DBS patient samples. The DBS samples of non-patient origin may not be suitable for assessing analytical accuracy of immunoassays.

基金:
语种:
被引次数:
WOS:
PubmedID:
中科院(CAS)分区:
出版当年[2018]版:
大类 | 3 区 化学
小类 | 2 区 分析化学 3 区 生化研究方法
最新[2025]版:
大类 | 2 区 化学
小类 | 3 区 生化研究方法 3 区 分析化学
JCR分区:
出版当年[2017]版:
Q1 CHEMISTRY, ANALYTICAL Q2 BIOCHEMICAL RESEARCH METHODS
最新[2023]版:
Q1 BIOCHEMICAL RESEARCH METHODS Q1 CHEMISTRY, ANALYTICAL

影响因子: 最新[2023版] 最新五年平均 出版当年[2017版] 出版当年五年平均 出版前一年[2016版] 出版后一年[2018版]

第一作者:
第一作者机构: [1]Department of Laboratory Medicine, The Second Affiliated Hospital of Guangzhou University of Chinese Medicine, Guangzhou, Guangdong, China
共同第一作者:
通讯作者:
通讯机构: [2]New York State Department of Health, Wadsworth Center, Empire State Plaza, Albany, NY, USA [3]College of Arts and Sciences, University at Albany, State University of New York, Albany, NY, USA
推荐引用方式(GB/T 7714):
APA:
MLA:

资源点击量:2018 今日访问量:0 总访问量:645 更新日期:2024-07-01 建议使用谷歌、火狐浏览器 常见问题

版权所有©2020 广东省中医院 技术支持:重庆聚合科技有限公司 地址:广州市越秀区大德路111号