Evaluation of a bracketing calibration-based isotope dilution liquid chromatography-tandem mass spectrometry candidate reference measurement procedure for 17α-hydroxyprogesterone in human plasma
机构:[1]Department of Laboratory Medicine, The Second Affiliated Hospital of Guangzhou University of Chinese Medicine, Guangzhou 510120, Guangdong, China大德路总院检验科大德路总院检验科广东省中医院[2]Second Clinical Medical College, Guangzhou University of Chinese Medicine, Guangzhou 510120, Guangdong, China广东省中医院[3]Department of Pharmacy and Guangdong Province Key Laboratory of Pharmacodynamic Constituents of Traditional Chinese Medicine & New Drug Research, Jinan University, Guangzhou 510632, Guangdong, China
A candidate reference measurement procedure (RMP) based on isotope dilution coupled with liquid chromatography-tandem mass spectrometry (ID-LC-MS/MS) was developed and validated for the quantification of 17 alpha-hydroxyprogesterone (17-OHP) in human plasma. 17-OHP spiked with a deuterium-labeled internal standard was extracted from plasma by liquid-liquid extraction with 1 mL n-hexane/ethyl acetate (3:2, v/v). Reversed-phase chromatography and positive electrospray ionization were used in the ID-LC-MS/MS. Gradient elution coupled with use of a C-18-packed ultrahigh-performance liquid chromatography column allowed complete baseline resolution of 17-OHP from its structural analogue desoxycorticosterone in 6 min. To determine the 17-OHP level in human plasma, a bracketing calibration method was used to give higher accuracy and precision. The limit of detection and the lower limit of the measuring interval for the candidate RMP were 2.1 pg/mL (6.4 pmol/L) and 4.6 pg/mL (13.9 pmol/L), respectively. Extraction recovery was determined to be (96.08 +/- 3.03)% (n = 3). Imprecision (intra-assay and interassay) was 4.03% or less at 0.83, 15.19, 64.22, and 313.46 ng/mL (2.51, 45.97, 194.34, and 948.56 nmol/L, respectively). Recoveries ranged from 98.05% to 102.24%. When comparing our RMP results with those obtained with an established RMP via International Federation of Clinical Chemistry and Laboratory Medicine external quality assessment scheme for reference laboratories in laboratory medicine (RELA) samples, we found that the biases ranged from -1.99% to 3.08% against the targets. No interference was observed, and the linear response ranged from 0.47 to 958.63 ng/mL (1.42 to 2900.90 nmol/L). Moreover, the candidate RMP was used to measure the concentration of 17-OHP in human plasma and was compared with an immunoassay using 40 plasma samples. The performance of the method meets the needs of an RMP (total coefficient of variation of 5% or less and bias of 3.08% or less). This method can be used for reference material value assignment of 17-OHP in human plasma matrix. It could also serve as an accurate reference baseline for routine methods to increase the accuracy and precision of certain clinical laboratory measurements.
基金:
National
Key Research and Development Program of China (2017YFF0205401),
the National Natural Science Foundation of China (81572088), the
Natur a l Sc i e n c e Fo u n d a t ion o f Guangdong Province
(2018A0303130124), the Specific Research Fund for TCM Science and
Technology of Guangdong Provincial Hospital of Chinese Medicine
(YN2016QJ15 and YN2019QL01), and the Guangzhou Science and
Technology Project (201704020213).
语种:
外文
被引次数:
WOS:
PubmedID:
中科院(CAS)分区:
出版当年[2018]版:
大类|3 区化学
小类|2 区分析化学3 区生化研究方法
最新[2025]版:
大类|2 区化学
小类|3 区生化研究方法3 区分析化学
JCR分区:
出版当年[2017]版:
Q1CHEMISTRY, ANALYTICALQ2BIOCHEMICAL RESEARCH METHODS
最新[2024]版:
Q1BIOCHEMICAL RESEARCH METHODSQ2CHEMISTRY, ANALYTICAL
第一作者机构:[1]Department of Laboratory Medicine, The Second Affiliated Hospital of Guangzhou University of Chinese Medicine, Guangzhou 510120, Guangdong, China[2]Second Clinical Medical College, Guangzhou University of Chinese Medicine, Guangzhou 510120, Guangdong, China
通讯作者:
推荐引用方式(GB/T 7714):
Zhang Qiaoxuan,Zhang Lu,Lin Haibiao,et al.Evaluation of a bracketing calibration-based isotope dilution liquid chromatography-tandem mass spectrometry candidate reference measurement procedure for 17α-hydroxyprogesterone in human plasma[J].ANALYTICAL AND BIOANALYTICAL CHEMISTRY.2019,411(27):7095-7104.doi:10.1007/s00216-019-02086-5.
APA:
Zhang, Qiaoxuan,Zhang, Lu,Lin, Haibiao,Cai, Zhiliang,Yan, Jun...&Huang, Xianzhang.(2019).Evaluation of a bracketing calibration-based isotope dilution liquid chromatography-tandem mass spectrometry candidate reference measurement procedure for 17α-hydroxyprogesterone in human plasma.ANALYTICAL AND BIOANALYTICAL CHEMISTRY,411,(27)
MLA:
Zhang, Qiaoxuan,et al."Evaluation of a bracketing calibration-based isotope dilution liquid chromatography-tandem mass spectrometry candidate reference measurement procedure for 17α-hydroxyprogesterone in human plasma".ANALYTICAL AND BIOANALYTICAL CHEMISTRY 411..27(2019):7095-7104