Simultaneous quantitation of endogenous estrone, 17β-estradiol, and estriol in human serum by isotope-dilution liquid chromatography-tandem mass spectrometry for clinical laboratory applications

文献详情

资源类型:
WOS体系:
Pubmed体系:

收录情况: ◇ SCIE ◇ EI

机构: [1]Department of Laboratory Medicine, The Second Affiliated Hospital of Guangzhou University of Chinese Medicine (Guangdong Provincial Hospital of Chinese Medicine), No. 111, Dade Road, Yuexiu District, Guangzhou, Guangdong 510120, China
出处:
ISSN:

关键词: Endogenous estrogens Mass spectrometry Liquid-liquid extraction Serumsamples Clinical laboratory

摘要:
Estrogen measurements are important in the assessment of female reproductive function and have expanding roles in other fields. A simple, accurate, highly sensitive and specific isotope-dilution liquid chromatography- tandem mass spectrometry method was developed and evaluated to simultaneously measure three endogenous estrogens in serum: estrone (E1), 17 beta-estradiol (E2), and estriol (E3). Chromatographic separation was achieved on aC18 column before electrospray ionization triple-quadrupole mass spectrometry in multiple reaction monitoring mode. The sample preparation in this assay requires no derivatization and extraction by liquid-liquid extraction. After optimization of the extraction conditions, the final extraction efficiency of E1, E2, and E3 was 83.8%, 78.9%, and 77.3% respectively. The metabolites and structural analogs that have the same molecular masses as the estrogens were separated under the optimized liquid chromatography conditions. Method validation showed satisfactory linearity over the concentration range of 20-10000 pg mL(-1) for all three estrogens (r(2) > 0.997). The limits of quantification were 5, 10, and 10 pg mL(-1) for E1, E2, and E3 respectively, and their recoveries ranged from 94.7% to 103.5%. The accuracy of the proposed method was further evaluated with use of certified reference materials BCR-576, BCR-577, and BCR-578 for E2 and 2014 International Federation of Clinical Chemistry and Laboratory Medicine External Quality Assessment Scheme for Reference Laboratories in Laboratory Medicine samples for E3, whose certified values were determined by reference methods. Great agreement was observed between the measured values and the certified values. Satisfactory precision (coefficients of variation less than 7.44%) was also obtained for the three estrogens. Moreover, the proposed method was successfully applied to measure the three estrogens in serum samples of pregnant women in the second trimester and to assess the accuracy of chemiluminescent immunoassays in clinical laboratories by determination of E2 and unconjugated E3 in serum samples.

基金:
语种:
被引次数:
WOS:
PubmedID:
中科院(CAS)分区:
出版当年[2016]版:
大类 | 3 区 化学
小类 | 2 区 分析化学 3 区 生化研究方法
最新[2025]版:
大类 | 2 区 化学
小类 | 3 区 生化研究方法 3 区 分析化学
JCR分区:
出版当年[2015]版:
Q1 CHEMISTRY, ANALYTICAL Q2 BIOCHEMICAL RESEARCH METHODS
最新[2023]版:
Q1 BIOCHEMICAL RESEARCH METHODS Q1 CHEMISTRY, ANALYTICAL

影响因子: 最新[2023版] 最新五年平均 出版当年[2015版] 出版当年五年平均 出版前一年[2014版] 出版后一年[2016版]

第一作者:
第一作者机构: [1]Department of Laboratory Medicine, The Second Affiliated Hospital of Guangzhou University of Chinese Medicine (Guangdong Provincial Hospital of Chinese Medicine), No. 111, Dade Road, Yuexiu District, Guangzhou, Guangdong 510120, China
通讯作者:
推荐引用方式(GB/T 7714):
APA:
MLA:

资源点击量:2022 今日访问量:0 总访问量:648 更新日期:2024-07-01 建议使用谷歌、火狐浏览器 常见问题

版权所有©2020 广东省中医院 技术支持:重庆聚合科技有限公司 地址:广州市越秀区大德路111号