Confirmed value of shear wave elastography for ultrasound characterization of breast masses using a conservative approach in Chinese women: a large-size prospective multicenter trial
机构:[1]State Key Laboratory of Oncology in SouthernChina, Department of Ultrasound, Sun Yat-SenUniversity Cancer Center, Guangzhou 510060, P. R.China[2]Department of Ultrasound, Fudan UniversityShanghai Cancer Center, Shanghai 200032, P. R. China[3]Department of Oncology, Shanghai Medical College,Fudan University, Shanghai 200032, P. R. China[4]Department of Ultrasound, First Affiliated Hospitalof Harbin Medical University, Harbin 150001, P. R.China[5]Department of Ultrasound, Sichuan Academyof Medical Sciences, Sichuan Provincial People’sHospital, Chengdu 610072, P. R. China四川省人民医院[6]Departmentof Ultrasound, West China Hospital of SichuanUniversity, Chengdu 610041, P. R. China[7]Departmentof Ultrasound, Sun Yat-Sen Memorial Hospital, SunYat-Sen University, Guangzhou 510235, P. R. China中山大学附属第二医院[8]Department of Ultrasound, Fujian Provincial CancerHospital, Fuzhou 350014, P. R. China[9]Departmentof Ultrasound, Shengjing Hospital of China MedicalUniversity, Shenyang 110004, P. R. China中国医科大学附属盛京医院[10]Departmentof Ultrasound, The Fourth Affiliated Hospital of ChinaMedical University, Shenyang 110032, P. R. China[11]Department of Ultrasound, Yinzhou Hospital Affiliatedto Medical School of Ningbo University, Ningbo315040, P. R. China[12]Department of Ultrasound,Binzhou Medical University Hospital, Binzhou 256603, P.R. China[13]Department of Ultrasound, Jiangsu ProvinceHospital, Nanjing 210029, P. R. China[14]Departmentof Ultrasound, The First Hospital of China MedicalUniversity, Shenyang 110001, P. R. China[15]Departmentof Ultrasound, Beijing Friendship Hospital of CapitalMedical University, Beijing 100050, P. R. China首都医科大学附属北京友谊医院[16]Department of Ultrasound, Guangdong ProvincialHospital of Traditional Chinese Medicine, Guangzhou510120, P. R. China[17]Department of Ultrasound, FirstAffiliated Hospital of PLA General Hospital, Beijing100037, P. R. China[18]Supersonic Imagine, Aix-en-Provence, France
Purpose: This study aimed to investigate the value of shear wave elastography (SWE) for characterization of breast masses in a Chinese population. Patients and methods: Two thousand two hundred seventy-three women consented to be prospectively enrolled for the characterization of breast masses with ultrasound and SWE. Breast masses were known from symptoms, palpability, and/or previous imaging screening with mammography and/or ultrasound. Correlation of SWE qualitative and quantitative features with malignancy risk and impact on diagnostic performance of combining SWE features were assessed, and the Breast Imaging Reporting and Data System (BI-RADS) scoring was calculated using histopathology as reference. Results: Data of 2,262 masses (median size: 13 mm; range: 1.3-50) from 2,262 patients (median age: 43 years; range: 18-91) were investigated, of which 752 (33.3%) were malignant. Sensitivity and specificity of BI-RADS diagnosis were 97.5% (733/752) and 54.8% (827/1,510), respectively. By logistic regression, the combination of maximum elasticity (E-max) measurements with BI-RADS assessments increased the area under the receiver operating characteristic curve from 0.908 (95% CI: 0.896-0.920) to 0.954 (95% CI: 0.944-0.962). Using E-max of 30 kPa or lower to selectively downgrade BI-RADS 4a masses to follow-up, and E-max of 160 kPa or higher to selectively upgrade BI-RADS 3 lesions to biopsy, specificity significantly increased from 54.8% (827/1,510) to 66.1% (998/1,510) (P<0.001) while sensitivity decreased nonsignificantly from 97.5% (733/752) to 96.9% (729/752) (P=0.2891). Positive predictive value for biopsy recommendation increased from 51.7% (733/1,417) to 58.7% (729/1,241) (P<0.001). Conclusion: Adding SWE maximum stiffness to BI-RADS 3 and BI-RADS 4a breast masses in a Chinese population increased significantly the specificity of breast ultrasonography, without significant change in sensitivity.
第一作者机构:[1]State Key Laboratory of Oncology in SouthernChina, Department of Ultrasound, Sun Yat-SenUniversity Cancer Center, Guangzhou 510060, P. R.China
通讯作者:
通讯机构:[2]Department of Ultrasound, Fudan UniversityShanghai Cancer Center, Shanghai 200032, P. R. China[3]Department of Oncology, Shanghai Medical College,Fudan University, Shanghai 200032, P. R. China[*1]No. 270 Dong-An Road, Department of Ultrasound, Fudan University Shanghai Cancer Center, Shanghai 200032, P.R. China
推荐引用方式(GB/T 7714):
Lin Xi,Chang Cai,Wu Changjun,et al.Confirmed value of shear wave elastography for ultrasound characterization of breast masses using a conservative approach in Chinese women: a large-size prospective multicenter trial[J].CANCER MANAGEMENT AND RESEARCH.2018,10:4447-4458.doi:10.2147/CMAR.S174690.
APA:
Lin, Xi,Chang, Cai,Wu, Changjun,Chen, Qin,Peng, Yulan...&Chen, Yaling.(2018).Confirmed value of shear wave elastography for ultrasound characterization of breast masses using a conservative approach in Chinese women: a large-size prospective multicenter trial.CANCER MANAGEMENT AND RESEARCH,10,
MLA:
Lin, Xi,et al."Confirmed value of shear wave elastography for ultrasound characterization of breast masses using a conservative approach in Chinese women: a large-size prospective multicenter trial".CANCER MANAGEMENT AND RESEARCH 10.(2018):4447-4458