关键词:
Anti-β2-glycoprotein I antibodies
Anticardiolipin antibodies
Antiphospholipid syndrome
Chemiluminescent immu-noassay
Performance evaluation
摘要:
Background: The diagnosis of antiphospholipid syndrome (APS) relies predominantly on the laboratory measure-ment of antiphospholipid antibodies (aPLs). We attempt to verify the analytical performance of anticardiolipin antibodies (aCL) IgA/IgG/IgM and anti-β2-glycoprotein I antibodies (aβ2GPI) IgA/IgG/IgM on a high-throughput automated immunoassay platform. Methods: Limit of blank (LOB), limit of detection (LOD), imprecision, and linearity were calculated according to the corresponding Clinical and Laboratory Standards Institute (CLSI) guidelines protocols. The biological refer-ence intervals (RIs) were verified in healthy individuals. Results: The LoB of aCL IgA/IgG/IgM and aβ2GPI IgA/IgG/IgM were 0.000, 1.200, 0.200, and 0.400, 1.250, 0.100, respectively. The LoD were 0.093, 1.715, 0.337 and 0.547, 2.174, 0.185 CU, respectively. All the within-run CVs and total CVs were less than the criterion at 10%. The linear analysis showed a good correlation between the pre-dictive values and observed values with correlation coefficients greater than 0.99. Conclusions: The BIO-FLASH automated chemiluminescent analyzer performed well in measuring aPLs. © 2021 Verlag Klinisches Labor GmbH. All rights reserved.
WOS:
WOS:000648746300014
PubmedID:
33865256
中科院(CAS)分区:
出版当年[2020]版:
大类
|
4 区
医学
小类
|
4 区
医学实验技术
最新[2025]版:
大类
|
4 区
医学
小类
|
4 区
医学实验技术
JCR分区:
出版当年[2019]版:
Q4
MEDICAL LABORATORY TECHNOLOGY
最新[2023]版:
Q4
MEDICAL LABORATORY TECHNOLOGY
影响因子:
0.7
最新[2023版]
0.8
最新五年平均
0.94
出版当年[2019版]
0.955
出版当年五年平均
0.955
出版前一年[2018版]
1.138
出版后一年[2020版]
第一作者:
Liang, E.-Y
第一作者机构:
[1]Department of Laboratory Medicine, The Second Affiliated Hospital, Guangzhou University of Chinese Medicine, Guangzhou, China
推荐引用方式(GB/T 7714):
Liang E.-Y,Ke P.-F,Li G.-H,et al.Performance evaluation of antiphospholipid antibodies on inova quanta flash system[J].CLINICAL LABORATORY.2021,67(4):988-994.doi:10.7754/Clin.Lab.2020.200809.
APA:
Liang, E.-Y,Ke, P.-F,Li, G.-H,Huang, H,Huang, W.-J...&He, M.(2021).Performance evaluation of antiphospholipid antibodies on inova quanta flash system.CLINICAL LABORATORY,67,(4)
MLA:
Liang, E.-Y,et al."Performance evaluation of antiphospholipid antibodies on inova quanta flash system".CLINICAL LABORATORY 67..4(2021):988-994