Light-initiated chemiluminescent assay of 17β-estradiol metrological traceability system established by manufacturer according to ISO17511:2020 and basic performance evaluation performed by clinical end-users

文献详情

资源类型:
WOS体系:

收录情况: ◇ SCIE

机构: [1]Second Affiliated Hospital of Guangzhou University of Chinese Medicine, Guangzhou, China [2]The Second Hospital of Tianjin Medical University, Tianjin, China [3]Standardization & Performance Evaluation Laboratory, Chemclin Diagnostics Co. Ltd, Beijing, China [4]Department of Clinical Immunology, School of Medical Laboratory, Tianjin Medical University, Tianjin, China
出处:
ISSN:

关键词: 17 beta-estradiol metrological traceability performance evaluation

摘要:
Background: In order to ensure the accuracy of the product, we established 1st model of metrological traceability hierarchy for light-initiated chemiluminescent assay (LICA) of 17 beta-estradiol (E-2) at the manufacturer, based on International Organization for Standardization (ISO) 17511:2020. Moreover, we verified/validated the basic performance (such as matrix effect and long-term stability of end-user IVD MD calibrator, precision, linearity interval, accuracy/ trueness, and detection capability) at the clinical end-user. Methods: Human serum samples were used in this study. E-2 was detected by mass spectrometry (MS) and LICA. The metrological traceability of LICA for E-2 was established according to ISO 17511: 2020 standards, and pools of human samples were used as the m.3. secondary calibrator. Precision was validated according to Clinical and Laboratory Standards Institute (CLSI) EP05-A3. The linear interval was verified according to CLSI EP06-ED2. Comparison of accuracy and trueness of E-2 with MS and Roche according to CLSI EP09-A3. The detection capability was validated according to EP17-A2. Matrix effect and long-term stability evaluation of end-user IVD MD calibrator were carried out according to CLSI EP14-A2, EP25-A. Statistical software was used for data analyses. Results: The use of pools of human samples and fine adjusting calibrators ensured the accuracy of end-user test results. The metrological traceability of LICA for E-2 was established. It showed excellent precision, meeting the requirements of allowable imprecision (7.5%). The allowable deviation from linearity (ADL) of 5% was allowed to show a good linear interval (12.52-4167.25 pg/ml). The accuracy/ trueness was verified, and relative deviation in the medical decision level met the performance specification of 10.03% compared with MS or Roche. The validated limit of blank, limit of detection, and limit of quantitation of E-2 were 4.95 pg/ml, 8.93 pg/ml, and 9.88 pg/ml, respectively (the allowed imprecision is 20.00%). The interference rate of E2 ranged from -5.5% to 6.6%. Conclusion: LICA showed high sensitivity, high specificity, excellent precision, wide linearity interval, IVD MD calibrator has long-term stability, and no matrix effect. The metrological traceability of E-2 established by using pools of human samples as M.3. can deliver accuracy to the end-user IVD MD and show good consistency with MS and Roche.

基金:

基金编号: 20201161

语种:
WOS:
PubmedID:
中科院(CAS)分区:
出版当年[2021]版:
大类 | 4 区 医学
小类 | 4 区 医学实验技术
最新[2025]版:
大类 | 4 区 医学
小类 | 3 区 医学实验技术
JCR分区:
出版当年[2020]版:
Q3 MEDICAL LABORATORY TECHNOLOGY
最新[2023]版:
Q2 MEDICAL LABORATORY TECHNOLOGY

影响因子: 最新[2023版] 最新五年平均 出版当年[2020版] 出版当年五年平均 出版前一年[2019版] 出版后一年[2021版]

第一作者:
第一作者机构: [1]Second Affiliated Hospital of Guangzhou University of Chinese Medicine, Guangzhou, China
通讯作者:
通讯机构: [1]Second Affiliated Hospital of Guangzhou University of Chinese Medicine, Guangzhou, China [*1]Department of Laboratory Science, Second Affiliated Hospital of Guangzhou University of Chinese Medicine, 111 Dade Rd., Guangzhou, Guangdong 510120 China
推荐引用方式(GB/T 7714):
APA:
MLA:

资源点击量:2018 今日访问量:0 总访问量:645 更新日期:2024-07-01 建议使用谷歌、火狐浏览器 常见问题

版权所有©2020 广东省中医院 技术支持:重庆聚合科技有限公司 地址:广州市越秀区大德路111号