A comparative evaluation of the analytical performances of Capillarys 2 Flex Piercing, Tosoh HLC-723 G8, Premier Hb9210, and Roche Cobas c501 Tina-quant Gen 2 analyzers for HbA1c determination

文献详情

资源类型:
WOS体系:
Pubmed体系:

收录情况: ◇ SCIE

机构: [1]Department of Laboratory Science, Second Affiliated Hospital of Guangzhou University of Chinese Medicine, Guangdong, China
出处:
ISSN:

关键词: Haemoglobin A(1c) capillary electrophoresis (CE) high-performance liquid chromatography (HPLC) affinity chromatography immunoassay

摘要:
Introduction: Haemoglobin A(1c) (HbA(1c)) is widely used in the management of diabetes. Therefore, the reliability and comparability among different analytical methods for its detection have become very important. Materials and methods: A comparative evaluation of the analytical performances (precision, linearity, accuracy, method comparison, and interferences including bilirubin, triglyceride, cholesterol, labile HbA(1c) (LA(1c)), vitamin C, aspirin, fetal haemoglobin (HbF), and haemoglobin E (Hb E)) were performed on Capillarys 2 Flex Piercing (Capillarys 2FP) (Sebia, France), Tosoh HLC-723 G8 (Tosoh G8) (Tosoh, Japan), Premier Hb9210 (Trinity Biotech, Ireland) and Roche Cobas c501 (Roche c501) (Roche Diagnostics, Germany). Results: A good precision was shown at both low and high HbA(1c) levels on all four systems, with all individual CVs below 2% (IFCC units) or 1.5% (NGSP units). Linearity analysis for each analyzer had achieved a good correlation coefficient (R-2 > 0.99) over the entire range tested. The analytical bias of the four systems against the IFCC targets was less than 6% (NGSP units), indicating a good accuracy. Method comparison showed a great correlation and agreement between methods. Very high levels of triglycerides and cholesterol (>= 15.28 and >= 8.72 mmol/L, respectively) led to falsely low HbA(1c) concentrations on Roche c501. Elevated HbF induced false HbA(1c) detection on Capillarys 2FP (> 10%), Tosoh G8 (> 30%), Premier Hb9210 (> 15%), and Roche c501(> 5%). On Tosoh G8, HbE induced an extra peak on chromatogram, and significantly lower results were reported. Conclusions: The four HbA(1c) methods commonly used with commercial analyzers showed a good reliability and comparability, although some interference may falsely alter the result.

基金:

基金编号: 2012BAI37B01 2012ZX09303009-003 81572088 2015A030313340 2014KT1593

语种:
被引次数:
WOS:
PubmedID:
中科院(CAS)分区:
出版当年[2015]版:
大类 | 3 区 医学
小类 | 3 区 医学实验技术
最新[2025]版:
大类 | 4 区 医学
小类 | 4 区 医学实验技术
JCR分区:
出版当年[2014]版:
Q1 MEDICAL LABORATORY TECHNOLOGY
最新[2023]版:
Q1 MEDICAL LABORATORY TECHNOLOGY

影响因子: 最新[2023版] 最新五年平均 出版当年[2014版] 出版当年五年平均 出版前一年[2013版] 出版后一年[2015版]

第一作者:
第一作者机构: [1]Department of Laboratory Science, Second Affiliated Hospital of Guangzhou University of Chinese Medicine, Guangdong, China
通讯作者:
推荐引用方式(GB/T 7714):
APA:
MLA:

资源点击量:2018 今日访问量:0 总访问量:645 更新日期:2024-07-01 建议使用谷歌、火狐浏览器 常见问题

版权所有©2020 广东省中医院 技术支持:重庆聚合科技有限公司 地址:广州市越秀区大德路111号